Instructions
Table Of Contents
MPI-BA-e-1710 
8 
9  Servicing, maintenance, disposal 
9.1  Cleaning 
Do not use aggressive cleaning agents (solvents or similar) but only a damp cloth 
with mild soapsuds, or household cleaner. Ensure that no liquid gets inside the unit 
and rub with a soft, dry cloth. 
Loose dirt can be removed carefully with a brush or hand-held vacuum cleaner. 
Balance must not be tilted or rotated for cleaning, as damage may occur. 
Remove dirt immediately. 
9.2  Servicing, maintenance 
The appliance may only be opened by trained service technicians who are authorized 
by KERN. 
9.3  Disposal 
Disposal of packaging and appliance must be carried out by operator according to 
valid national or regional law of the location where the appliance is used. 
10 Error message 
Display  Description / help 
Err0  Zero range exceeded 
•  Unload weighing plate 
•  Tap weighing platform slightly and “0000“ will appear. 
•  The scales will turn off automatically. 
•  Repeat the weighing process 
Err1  Stabilisation error 
•  Repeat the weighing process 
•  Remain as still as possible during the weighing process. 
Err2  Overload 
•  Unload weighing platform immediately, risk of damage 
to scales 