Datasheet
PROPRIETARY INFORMATION:
PROPRIETARY OF GLOBTEK, INC. ANY REPRODUCTION, DISCLOSURE OR
USE OF THIS DRAWING, IN WHOLE OR IN PART, IS HEREBY
PROHIBITED EXCEPT AS SPECIFIED IN WRITING BY GLOBTEK, INC.
REVISION
REV
A
DESCRIPTION
INITIAL RELEASE, RFS.23008, CL
DATE
05/25/12
APPROVED
HM
TABULATION BLOCK
TOLERANCES:
FSCN No.:
G, N
SIZE
A
MODEL NO: GTM41076-0612
REV.
A
CAGE CODE: 0GV V90
SCALE: NONE
PART NO: WR9QE500CCP-N-MED-WD
SHEET 1 OF 6
INIT.BY: DMR
DATE:
07/19/06
DRAWN: NF
DATE:
07/19/06
APRVD : HM
DATE:
07/19/06
PART NO REV.
DESCRIPTION
NOTES
www.globtek.com
DECIMALS
.xx +/- 0.1
.xxx +/- .005
ANGULAR
+/- 1
o
MILLIMETERS
.xx +/- 0.3
.xxx +/- .013
ANGULAR
+/- 1
o
1. NOTES:
DIMENSIONS ARE IN MM UNLESS SPECIFIED OTHERWISE.
2. ELECTRICAL SPECIFICATIONS:
INPUT VOLTAGE: 100-240 VAC
INPUT CURRENT: 0.5 A RMS MAX
INPUT FREQUENCY: 47 - 63Hz
OUTPUT VOLTAGE : 12VDC
OUTPUT CURRENT : 0.5A,
NO MINIMUM LOAD REQUIRED TO MAINTAIN OUTPUT VOLTAGE REGULATION
OUTPUT POWER (RATED): 6WATTS MAX
SWITCHING FREQUENCY
OUTPUT RIPPLE (PEAK TO PEAK): 1% OR 50 mV WHICHEVER IS GREATER. MEASURED AT 20 MHz
BANDWIDTH WITH 0.1 µf CERAMIC CAPACITOR IN PARALLEL WITH 10 µf
ELECTROLYTIC CAPACITOR CONNECTED AT THE END OF OUTPUT CONNECTOR
OUTPUT LOAD REGULATION: +
5 % OF NOMINAL OUTPUT VOLTAGE, MEASURED AT O/P CONNECTOR AT
NOMINAL LINE
LINE REGULATION: +
1 % OF NOMINAL OUTPUT VOLTAGE MEASURED AT O/P CONNECTOR
TURN ON/ TURN OFF OVERSHOOT: 5% MAX 500uS MAX RECOVERY TIME FOR 25% STEP LOAD
TURN-ON DELAY: 3000 mSEC MAX
HOLD-UP TIME: 8 mSEC MIN AT NOMINAL INPUT AND FULL LOAD
INRUSH CURRENT: 30 A MAX AT COLD START 120 VAC, 60A MAX COLD START 230VAC
SWITCHING FREQUENCY: 66.5 KHz TYPICAL
OVER-VOLTAGE PROTECTION: PROTECTED, UNIT WILL AUTORECOVER UPON REMOVAL OF FAULT
OVER-CURRENT / SHORT CIRCUIT: PROTECTED, UNIT WILL AUTORECOVER UPON REMOVAL OF FAULT
OTHER PROTECTION: INPUT LINE AND NEUTRAL FUSING
SAFETY APPROVALS : UL60601-1, CUL TO 22.2 NO. 601.1-M90, INNOVA BAUART TO EN60601-1
CE CLASS II, PSE TO J60601-1, CB REPORT TO IEC 60601-1:2005 (3rd EDITION)
AS/NZ TO 60601-1, C-TICK TO EN55011, CCC
DIELECTRIC WITHSTAND VOLTAGE: 5656 VDC PRIMARY-SECONDARY
LEAKAGE CURRENT: <
100 µA AT 240 VAC
EMI: COMPLIES WITH EN55011 CLASS B AND FCC PART 15 CLASS B
WHEN TESTED WITH A RESISTIVE LOAD, BOTH CONDUCTED AND RADIATED,VCCI
CE MARK: TESTED TO COMPLY WITH EN61000-3-2, EN61000-3-3, EN60601-1-2 2001,
INCLUDING EN61000-4-2, EN61000-4-3, EN61000-4-4, EN61000-4-5,
EN61000-4-6, EN61000-4-8, EN61000-4-11.
EFFICIENCY: COMPLIES TO SECTION 301 OF ENERGY INDEPENDENCE AND SECURITY ACT (EISA)
COMPLIES WITH ENERGY STAR TIER 2 (NORTH AMERICA),
CECP TIER 2 (CHINA), MEPS TIER 2 (AUSTRALIA), CODE OF CONDUCT (EUROPE)
5.1 ~ 6.9V WITH 1000 mm CABLE ONLY
MTBF: 200,000 HOURS AT 25°C AMBIENT TEMPERATURE
OPERATING TEMPERATURE: 0°C TO 40° C
STORAGE TEMPERATURE: -40°C TO 80° C
HUMIDITY: 0% TO 90% RELATIVE HUMIDITY
ROHS: COMPLIES WITH EU 2002/95/EC AND CHINA SJ/T 11363-2006
Limitation of Use:
Globtek product are not authorized for use as
mission critical components in life support,
hazardous environment, nuclear or
aircraft applications without prior written
approval from the CEO of Globtek Inc.
Contents of this document are subject to
change without prior notice
Foot Note:
Globtek Inc. will not be liable for the
safety and performance of these power supplies
if unauthorized access and repair occurs. End
user should consult applicable UL, CSA or
EN standards for proper installation instructions
All models under this series are Certified to be in compliance with all the applicable requirements of UL60601-1 2nd Edition, CSA 22.2 No.601.1-M90 and EN 60601-1:A2. Globtek's Medical power
supplies are designed to be used in applications that do not connect the output to Patient Circuits directly. The power supply can be used in systems with patient applied parts only when appropriate
system level isolation is provided by the end user.
C
DWG TITLE
:
WALL PLUG-IN SWITCHING POWER SUPPLY, 6WATTS, 12V@ 0.5A
UNIVERSAL INPUT, MEDICAL, CLASS II