User manual
HFB-BA-e-1212 
24 
Overview: 
Function 
Available 
settings 
Description 
F0 CAP  
Select capacity 
30k  Max = 30 kg 
Modifications may only be carried out by  
a specialist with competent knowledge. 
60k  Max = 60 kg 
150k  Max = 150 kg  
Default setting HFB 150K50 
300k  Max = 300 kg  
Default setting HFB 300K100 
600k  Max = 600 kg  
Default setting HFB 600K200 
F1 unit  Not documented 
F2 res 
Select resolution 
inC 5*  5  
inC 10  10  
F3 a2n 
Automatic zero point 
correction 
(zero tracking) 
0.5 d  0.5 d 
1 d*  1 d 
2 d  2 d 
4 d  4 d 
F4 Flt 
Filter 
Flt 1  fast 
slow
Flt 2* 
Flt 3 
Flt 4 
Flt 5 
F5 inP  Internal A/D converter value 
F6 bk 
Display background 
illumination 
bk Au  Background illumination switches on 
automatically when loaded or a  
button is pressed 
bk of  Background illumination off 
bk on  Background illumination on 
F7 off 
Automatic shutdown 
function 
of 0 
Balance will switch-off automatically 
after the set period. 
Selectable 0/3/5/15/20 minutes. 
of 3 
of 5 
of 10 
of 20 
F8 Grv  Not documented 
F9 CAL  Adjustment, see chapt. 7 
* = default setting 