User manual
18    GAB-N-BA-e-1315 
9.2  Non-verifiable models: 
  Switch on balance Whilst balance is carrying out self-test 
(counts down from 99… to 00…) 
TARE
 and 
%
 press at the 
same time. 
“UnLoad“ will be displayed, followed by flashing numeric 
value. 
 
(Example) 
  Change by pressing the navigation keys (See manual chap. 
8); select desired setting; currently enabled digit will be 
flashing. 
(Example) 
  Confirm value by 
0
, “Load“ will be displayed. 
  Place adjustment weight, stability display   appears 
  After the adjustment the balance will carry out a self-test. 
Remove adjusting weight during self test, 
balance will return into weighing mode automatically. 
In case of an adjustment error or incorrect adjusting weight 
the display will show an error message, repeat adjustment 
process. 
GROSS
ZERO