8  Troubleshooting 
The balance should be switched off for a short time following an interruption in the 
programme sequence and disconnected from the mains supply. It is then necessary 
to repeat the weighing process from the beginning. 
Help: 
Interruption Possible cause 
Weight display is not illuminated. 
•  The balance is not switched on. 
•  The mains supply connection has been 
interrupted (mains cable not plugged 
in/faulty). 
•  Power supply interrupted. 
•  The recharchable battery pack is empty. 
The weight display changes continu-
ally 
• Draught/air movement 
•  Table/floor vibrations 
•  The weighing plate is in contact with for-
eign matter. 
•  Electromagnetic fields / static charging 
(choose different location/switch off inter-
fering device if possible) 
The weighing result is obviously in-
correct 
•  The balance display is not set to zero 
•  Adjustment is no longer correct. 
•  Great fluctuations in temperature. 
•  Electromagnetic fields / static charging 
(choose different location/switch off inter-
fering device if possible) 
Switch the balance off if other error messages should appear and then switch on 
again. Contact the manufacturer if the error message does not disappear. 