Procedure when adjusting: 
Observe stable environmental conditions. 
A warming up time (see chapter 1; Technical Data) is required for stabilization. 
Ensure that there are no objects on the weighing plate. 
Balance display 
Operation 
  Start balance by pressing   
The balance carries out an autotest and changes 
then into the weighing mode, the zero display „0.0“ 
appears 
(example) 
(example) 
  Press   for approx. 3 seconds, „CAL“ appears 
shortly, then „CAL 0“ and after that „LoAd“ is 
displayed alternately with the value of the first 
adjustment weight. 
  Place the first adjustment weight in the middle of the 
weighing plate. 
After a short time the value of the second 
adjustment weight is displayed alternately with 
„LoAd“ 
  Remove the the first adjustment weight from the 
weighing plate and place the second adjustment 
weight in the middle of the weighing plate. 
Adjustment occurs automatically after standstill 
control. 
„donE“ will be shortly displayed. Then the value of 
the second adjustment weight will be displayed. 
Adjustment process has been finished successfully. 
The balance is now in weighing mode. 