User manual
Overview of the possible software settings: 
(there may be small differences referring to the model): 
         Navigation Menu 
GROSS
NET
 0
Press  TARE 3sec
SET
End
md-2md-1
15.010.0
6.05.0
3.02.0 3.02.0
1.51.0
0.0020.0050.01
1g
2g5g
0.5g1g2g
0.2g0.5g1g
19200960048002400
OffOn
OffOn
trAdtrAtEtrAnGrAvdivSA_OFF
CAPCAP
Overload setting by
hardware option
Off300240180120
If the TRAD setting is OFF, this must be set to ON before a verification. 
Reset into verifiable setting: 
Press the 
 key and keep it pressed until „A_OFF“ appears on the display. 
Press 
 key repeatedly until „TRAD“ appears. 
Now press 
 key, in the display appears “OFF”. 
Now press the 
 key, „ON“ appears. 
Then press subsequently 
,   ,   . 
The balance is now ready for verification. 
In order to check the settings, repeat the process. 
FOB-BA-e-1118 