5.7.3  Non-verifiable Models 
FOB 1.5K0.5, FOB 3K1, FOB 6K2, FOB 7.5K0.5L, FOB 15K1L, FOB 30K2L 
Ö  Turn balance off and hold   key, test procedure appears on the display, op-
erate the   -key 3x times during this test procedure. 
Ö  Now let go of both keys. In the display appears “CAL”. Press key   in dis-
play appears “Load”. 
Ö  Place the tare weight (nominal value, see “Techn. Specifications“ chapter 1) in 
the centre of the weighing pan, then press the key  . “CAL” will appear on 
the display for approx. 2 seconds, followed by the nominal value of the adjust-
ing weight. 
Ö  Remove the adjusting weight from the weighing plate to complete adjusting. 
5.7.4  Check software settings for verification of a balance 
(only models FOB 1K1M, FOB 2K2M, FOB 5K5M): 
For the adjustment function, the balance must be switched over to a service 
mode. This service mode is activated by actuating the adjustment switch 
twice. 
In the service mode all parameters of the balance can be modified. There is a 
special setting named „TRAD“. 
The service parameters may not be modified, as this could damage the bal-
ance settings. 
Especially the setting „TRAD“ is very important for a verification. This setting 
activates and deactivates the verifiable settings of the balance. 
Prior to a verification the setting „TRAD“ must be revised. 
Press the 
 key and keep it pressed until „A_OFF“ appears on the display. 
Press 
 button repeatedly. The setting „Trad“ must not appear (see 
graphic). 
FOB-BA-e-1118 
13