User manual
FFN-N-BA-e-1222    20 
7.1.2  Adjusting verifiable models 
The adjustment is locked for verified balances. 
with adjustment switch 
To disable the access lock, destroy the seal and actuate the adjustment 
switch. Position of the adjustment switch see chapter 7.2.1 
Attention: 
After destruction of the seal the balance must be re-verified by an 
authorised agency and a new verification wire/seal mark fitted before it can 
be reused for applications subject to verification. 
Display 
Operation 
  Start balance by pressing   
  While the balance carries out a self test (00…- 99…), press 
 and keep pressed until „F1 CAL“ appears on the 
display. 
  Actuate the adjustment switch on the lower side of the 
balance 
   Press   
in the display appears „UnLod“. 
(Beispiel) 
   Press   anew 
The weighing value is displayed. 
  Press   to enter the value of the adjustment weight 
(see chapter 1) 
  Place the adjustment weight 
   Press   
While the balance carries out a self-test, take away the 
adjustment weight 
The balance changes over to zero display. 
Now the process of adjustment is completed. 