3  Basic Information (General) 
3.1  Proper use 
The balance you purchased is intended to determine the weighing value of material 
to be weighed. It is intended to be used as a “non-automatic“ balance, i.e. the 
material to be weighed is manually and carefully placed in the centre of the weighing 
plate. As soon as a stable weighing value is reached the weighing value can be read. 
3.2  Improper Use 
Do not use balance for dynamic weighing. In the event that small quantities are 
removed or added to the material to be weighed, incorrect weighing results can be 
displayed due to the “stability compensation“ in the balance. (Example: Slowly 
draining fluids from a container on the balance). Do not leave permanent load on the 
weighing plate. This may damage the measuring system. Impacts and overloading 
exceeding the stated maximum load (max) of the balance, minus a possibly existing 
tare load, must be strictly avoided. Balance may be damage by this. 
Never operate balance in explosive environment. The serial version is not explosion 
protected. 
The structure of the balance may not be modified. This may lead to incorrect 
weighing results, safety-related faults and destruction of the balance. 
The balance may only be used according to the described conditions. Other areas of 
use must be released by KERN in writing. 
3.3  Warranty 
Warranty claims shall be voided in case 
•  Our conditions in the operation manual are ignored 
•  The appliance is used outside the described uses 
•  The appliance is modified or opened 
•  Mechanical damage and damage caused by media, liquids 
•  Natural wear and tear 
•  The appliance is improperly set up or incorrectly electrically connected 
•  The measuring system is overloaded 
3.4  Monitoring of Test Resources 
In the framework of quality assurance the measuring-related properties of the 
balance and, if applicable, the testing weight, must be checked regularly. The 
responsible user must define a suitable interval as well as type and scope of this test. 
Information is available on KERN’s home page (www.kern-sohn.com) with regard to 
the monitoring of balance test substances and the test weights required for this. In 
KERN’s accredited DKD calibration laboratory test weights and balances may be 
calibrated (return to the national standard) fast and at moderate cost. 