User manual
rE Cr: 
The remote control commands s/w/t are sent from the remote control unit to the 
balance as ASCII code. After the balance having received the s/w/t commands, it will 
send the following data. 
Take into account that the following remote control commands must be sent 
without a subsequent CR LF. 
s 
Function:  Stable weighing value for the weight is sent via the RS232 
interface 
w 
Function:  Weighing value for the weight (stable or unstable) is sent 
via the RS232 interface 
t 
Function:  No data are sent, the balance carries out the tare function. 
h.  Format for stable values for weight/quantity/percentage 
1  2  3  4  5  6  7  8  9  10 11 12 13 14 15 16 17 18
B
* 
B B B B B B B B 0 .  0 B g B B CR LF
i.  Format in case of fault 
1 2 3 4 5 6 7 8 9 10 11 12 13 14 
B B B B B B B E r  r  o  r  CR LF 
j.  Format for unstable values for weight/quantity/percentage 
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 
B
* 
B B B B B B B B 0 .  0 B CR LF 
Symbols 
B
*
  Blank or M 
B / 0 /.  Blank / weight values / decimal dot, depending on weighing value 
g  Weight unit / pcs / % 
E, o, r  ASCII code or “E, o, r” 
CR  Carriage return symbol 
LF  Line feed symbol 
FCB-BA-e-0911 