User manual
EW/EG-N/EWB-e-1226 
54 
8.5  Data input commands 
8.5.1  Command input format 
is made up of 4 characters, CR=0DH, LF=0AH 
1 
2 
3 
4 
C1 
C2 
CR 
LF 
8.5.2  External taring command 
C1 
C2 
Code 
Content 
T 
SP 
54H 
20H 
Tare out command 
8.5.3  External control commands 
C1 
C2 
Code 
Meaning  
O 
0 
4FH 
30H 
No data output  
O 
1 
4FH 
31H 
Continuous data output  
O 
2 
4FH 
32H 
Continuous output of stable weight value data  
O 
3 
4FH 
33H 
Output of stable and unstable weight values after pressing the 
PRINT button  
O 
4 
4FH 
34H 
Data output upon stable weight value, having previously unloaded 
the weighing scale  
O 
5 
4FH 
35H 
Data output upon stable weight value. No data output at unstable 
weight value. Data output again, after stabilization has been 
reached 
O 
6 
4FH 
36H 
Data output upon stable weight value. Continuous data outputting 
at unstable weight value. 
O 
7 
4FH 
37H 
Output of stable weight values after pressing the PRINT button  
O 
8 
4FH 
38H 
One-time immediate data output* 
O 
9 
4FH 
39H 
One-time data output after stabilization has been reached* 
O 
A 
4FH 
41H 
One-time, immediate data output after lapse of a defined period of 
time* 
O 
B 
4FH 
42H 
One-time, immediate data output after lapse of a defined period of 
time and upon stable weight value* 
* While using these commands for remote control purposes, do not press the PRINT button (data 
transmission error). In case data transmission error occurs, briefly cut off the weighing scale from its 