User manual
EW/EG-N/EWB-e-1226
38
6.2.3 Percent weighing
(excl. model KERN EW120-4NM)
Display symbol: %
Percent weighing allows the weight to be displayed as a percentage in relation to a
reference weight. The displayed weight value is adopted as a fixed prescribed per-
cent value. (Standard setting: 100%).
Operation
Display
1. Use the key to switch the bal-
ance on. Use the key to select unit
change-over [ % ] (see chap. 6.2.1).
Information:
Tare containers can also be used during
percent weighing. Before starting percent
weighing use the key to tare out the
container.
2. Operate the key.
[P. SEt] will flash on the display.
3. Place the reference weight = 100% in
the balance dish.
4. Operate the key.
The reference weight is stored.
ο 0 I I I F
0 %
P. SEt %
P. SEt %
ο 0 I I I F
100.00 %