User manual
EW/EG-N/EWB-e-1226 
26 
5.10  Verification switch and official seal 
Prior to verification the verification switch must be moved from the displayed position 
(see direction of arrow) to verification position. Once in this position, a parenthesis 
will be displayed around the last display point.  
Following calibration the balance is sealed at the marked position.  
Balance calibration is not valid without the “official seals“.  
Position of the “official seals“: 
                 Move verification switch 
Balances that are subject to compulsory verification must be taken out of op-
eration if: 
  The weighing result of the balance is outside the error limit. Therefore, apply a 
known test weight (approx. 1/3 of the nominal load) to the balance at regular in-
tervals and compare with the display value.  
  The reverification deadline has been exceeded. 
Official seal 
Official seal 