User manual
EW/EG-N/EWB-e-1226 
38 
6.2.3  Percent weighing 
(excl. model KERN EW120-4NM) 
Display symbol: % 
Percent weighing allows the weight to be displayed as a percentage in relation to a 
reference weight. The displayed weight value is adopted as a fixed prescribed per-
cent value. (Standard setting: 100%). 
Operation 
Display 
1.  Use the   key to switch the bal-
ance on. Use the   key to select unit 
change-over [ % ]  (see chap. 6.2.1). 
Information: 
Tare containers can also be used during 
percent weighing. Before starting percent 
weighing use the   key to tare out the 
container. 
2.  Operate the   key. 
[P. SEt] will flash on the display. 
3.  Place the reference weight = 100% in 
the balance dish. 
4.  Operate the   key. 
The reference weight is stored.  
 ο    0 I I I F   
       0 % 
   P. SEt  % 
   P. SEt  % 
 ο    0 I I I F   
    100.00 % 