User manual
14        EMS-BA-e-1213 
  Weigh the material, the net weight will be indicated. 
246
g
The taring process can be repeated any number of times, 
e.g. when adding several components for a mixture 
(adding). The limit is reached when the whole weighing 
range is exhausted. 
The weight of the weighing container will be displayed as a 
minus number after removing the weighing container. 
The tare weight is saved until it is deleted. 
Delete tare 
  Unload the balance and press the TARE button, the zero 
display appears. 
0
g
PRE-TARE 
function 
Using this function the weight of a tare vessel is stored. Even 
after turning off/on the weighing balance will continue working 
with the saved tare value. 
  In weighing mode put tare vessel on the weighing plate 
  Press repeatedly the MODE key until „PtArE“ flashing 
appears. 
  Use SET key to store the current weight on the weighing 
plate as a PRE-TARE value. 
Delete PRE-TARE 
value 
  Remove all loads from the balance, press TARE and press 
repeatedly the MODE key until „PtArE“ flashing appears. 
  Use the SET key to confirm. The PRE-TARE value is 
deleted, the zero display appears. 