Instructions
Table Of Contents
EMB-S-BA-e-1612 
17 
8  Instant help 
In case of an error in the program process, briefly turn off the balance and disconnect 
from power supply. The weighing process must then be restarted from the beginning. 
Help: 
Fault  Possible cause 
The displayed weight does not glow. 
•  The balance is not switched on. 
•  Batteries are inserted incorrectly or 
empty 
•  No batteries inserted. 
•  The mains supply connection has 
been interrupted (mains cable not 
plugged in/faulty). 
•  Power supply interrupted. 
The displayed weight is permanently 
changing 
•  Draught/air movement 
•  Table/floor vibrations 
•  Weighing pan has contact with other 
objects. 
• 
Electromagnetic fields / static charging 
(choose different location/switch off 
interfering device if possible) 
The weighing result is obviously 
incorrect 
•  The display of the balance is not at 
zero 
•  Adjustment is no longer correct. 
•  Great fluctuations in temperature. 
• 
Electromagnetic fields / static charging 
(choose different location/switch off 
interfering device if possible) 
Should other error messages occur, switch balance off and then on again. If the error 