User manual

EW/EG-N/EWB-e-1226
47
7.2 List of the function parameters
The balance has been set to a certain standard configuration in the factory. This is
marked by a .
Function
Display
Choice
Description of the choice
possibilities
Bar graph
1
b.G.
0
Off
1
On
Tolerance weighing
(excl. EW-120-4NM)
2
SEL
0
Off
1
On (chap. 7.2.1)
Zero alignment
3
A.0
0
No zero point correction
1
Automatic zero point correction activated.
Automatic shutoff after 3 min.
for rechargeable battery op-
eration (function is only avail-
able inrechargeable battery
mode)
4
A.P.
0
Automatic shutoff deactivated for re-
chargeable battery operation (optional).
1
Automatic shutoff activated for recharge-
able battery operation (optional).
Display speed
5
rE.
0
Setting for metering
1
Sensitive and fast
2
3
4
5
Non-sensitive but slow
Vibratory filter
6
S.d.
1
Sensitive and fast (very tranquil installation
location).
2
3
4
Non-sensitive but slow (very unsettled
installation location).
5
only EW-120-4NM
6
only EW-120-4NM
Interface
7
I.F.
0
Interface not active
(excl. EWB models)
1
6-digit data format (excl. EW-120-4NM)
2
7-digit data format (chap. 7.2.2)
Weight unit
(only selectable, if the cali-
bration switch is not in the
calibration position see
Ch. 5.10)
81
85
S.u.
101
(g)
214
(ct)
15
(oz)
16
(lb)
17
(ozt)
18
(dwt)
19
(grain), (excl. EWB models)
1A
(tl Hong Kong)
1b
(tl Singapore,Malaysia)
1C
(tl Taiwan)
1d
(mom)
1E
(to)
320
(Pcs) excl. EW-120-4NM
4IF
(%) excl. EW-120-4NM
500
Unit not set (can not be set at 81.S.u.)