User manual
EW/EG-N/EWB-e-1226 
48 
Not documented 
9. 
Ai 
0 
Not documented 
☆1 
Always use this setting. 
Data-out 
(only selectable, if the cali-
bration switch is not in the 
calibration position see 
Ch. 5.10) 
A. 
PrF. 
1 
No printout possible, if the last display location 
is enclosed in brackets. 
☆2 
Printout possible, even if the last display loca-
tion is enclosed in brackets. 
Comment: Always select this setting, before 
the balance is calibrated, as this menu item 
can no longer be called due to the calibration 
3 
The printout is only carried out, if the calibration 
switch is not in the calibration position see 
Ch. 5.10. 
Enhanced log printout follow-
ing adjustment 
(only selectable on 
EG models) 
0 
GLP 
0 
Off 
☆1 
On 
**CALIBRATION** 
MODEL: 
S/N: 
ID: 
DATA: 
TIME: 
 *CAL. END 
NAME 
******************** 
 Title 
 Model of balance 
 Serial no. 
 ID no. 
 Date of calibration 
 Time of calibration 
 End of calibration 
 Signatur 