User manual
DE-BA-e-1255 
34 
8.3.4  rE Cr 
The remote control commands s/w/t are sent from the remote control unit to the 
balance as ASCII code. After the balance having received the s/w/t commands, it will 
send the following data. 
Take into account that the following remote control commands must be sent 
without a subsequent CR LF. 
s 
Function: 
Stable weighing value for the weight is sent via the RS232 
interface 
w 
Function: 
Weighing value for the weight (stable or unstable) is sent 
via the RS232 interface 
t 
Function: 
No data are sent, the balance carries out the tare function. 
h.  Format for stable values for weight/quantity/percentage 
1 
2 
3 
4 
5 
6 
7 
8 
9 
10 
11 
12 
13 
14 
15 
16 
17 
18 
M 
S 
N
1
N
2
N
3
N
4
N
5
N
6
N
7
N
8
N
9
N
10
B  
U
1
U
2
U
3
CR  
LF 
i.  Format in case of fault  
1 
2 
3 
4 
5 
6 
7 
8 
9 
10 
11 
12 
13 
14 
15 
16 
17 
18 
B  
B  
B  
B  
B  
B  
B  
B  
B  
B  
B  
E  
r  
r  
o  
r  
CR  
LF 
j.  Format for unstable values for weight/quantity/percentage 
1 
2 
3 
4 
5 
6 
7 
8 
9 
10 
11 
12 
13 
14 
15 
16 
17 
18 
M 
S 
N
1
N
2
N
3
N
4
N
5
N
6
N
7
N
8
N
9
N
10
B  
B  
B  
B  
CR  
LF 
Symbols 
M 
Blank or M 
S 
Blank or minus sign (-) 
N
1
 … N
10
10 numeric ASCII codes for weight values including decimal places or 
blanks 
U
1
 … U
3
3 ASCII codes for weighing unit pcs. / % / or blank 
B 
Blank 
E, o, r 
ASCII code or “E, o, r” 
CR 
Carriage Return 
LF 
Line Feed 