User manual
CPB-N / CPB-DM-BA-e-1323 
17 
6.7.2  Models CPB-DM (verified models) 
The adjustment is locked for verified balances. Carrying out adjustment 
requires that the seal is destroyed and the adjusting switch is pressed when 
turning on the scale. For position of adjusting switch, see chap. 6.9.1. 
Attention: 
After destruction of the seal the balance must be re-verified by an 
authorised agency and a new verification wire/seal mark fitted before it can 
be reused for applications subject to verification. 
Operation 
indikation 
  Switch-on balance and during the selftest 
actuate the adjustment switch and press 
the TARE button 
  Use the number keys to enter password: 
either 
  Default password "0000"  
or 
  Personal password, enter under 
function      see chap.12 
  Confirm with TARE button 
 