User manual
4 Density determination of solids
for the density determination of solid material, the solid is weighed first in air and then
in the measuring liquid. From the weight difference results the buoyancy from where
the software calculates the density.
4.1 Activate function
In the „Unit.SEL“ menu the density determination function for solid material „U-
T
d“
can be activated, which then will be available for the operator without needing to
enter the menu each time. The activated function then can be called-up directly via
the UNIT/ button.
Ö At the gram display, press the CAL/MENU button repeatedly
until „FUnC.SEL“ is displayed.
Ö Press the TARE/ button.
(Example)
Ö Press the CAL/MENU button repeatedly until „Unit.SEL“ will
be displayed.
Ö Press the TARE/ button.
(deactivated)
(activated)
Ö Press the CAL/MENU button repeatedly until „U-
T
d“ will
be displayed.
If the function of density determination already has been
activated, the standstill display ( ) will appear. In this case
press the ON/OFF button repeatedly, the balance returns
into the menu/weighing mode.
If the function for density determination of solid material is
deactivated, activate it by the TARE/
button.
The standstill display (
) will appear.
Press the ON/OFF button repeatedly, the balance returns
into the menu/weighing mode.
ABS-A02-BA-e-1010 9