9  Data output RS 232 C 
9.1  Technical data 
  8-bit ASCII Code 
  1 start bit, 8 data bits, 1 stop bit, no parity bit 
  Baud rate selectable at 1200, 2400, 4800 and 9600 Baud 
  Miniature plug-in necessary (9 pole D-Sub) 
  For operation with interface faultless operation is only ensured with the correct 
KERN – interface cable (max. 2m) 
9.2  Pin allocation of the balance output socket (front view), 
5     1 
 9 6 
Pin 2:  Transmit data 
Pin 3:  Receive data 
Pin 5:  Signal ground 
9.3  Explanation of the data transfer 
9.3.1  Pr PC 
Press the PRINT key, at stable value the format is transferred from LAPR. 
a.  Format for stable values for weight/quantity/percentage 
1  2  3  4  5  6  7  8  9  10 11 12 13 14 15 16 17 18
B
* 
B B B B B B B B 0 .  0 B g B B C
R 
LF
b.  Format in case of fault 
1 2 3 4 5 6 7 8 9 10 11 12 13 14 
B B B B B B B E r  r  o  r  CR LF 
9.3.2  AU Pr 
As soon as the weighing value is stable, the format is automatically transferred from 
LAPR. 
c.  Format for stable values for weight/quantity/percentage 
1  2  3  4  5  6  7  8  9  10 11 12 13 14 15 16 17 18
B
* 
B B B B B B B B 0 .  0 B g B B C
R 
LF
d.  Format in case of fault 
1 2 3 4 5 6 7 8 9 10 11 12 13 14 
B B B B B B B E r  r  o  r  CR LF 
440-BA-e-0941 
28 