User manual
8.5  Interface RS232C 
Data output via interface RS 232 C 
General Information 
The previous condition for the data transfer between balance and a peripheral device 
(e.g. printer, PC ...) is that the appliances are set to the same interface parameters 
(e.g. baud rate, transfer mode ...). 
8.5.1  Data transfer mode 
Setting of data transfer mode  Balance display 
1.   Keep the   key pressed until „Unit" is displayed. 
Unit 
2. Press   key, „Pr“ appears. 
Pr 
3.  To change the setting, press the   key. 
Pr PC 
4.  The mode can be converted using the   key 
(Pr PC; AU PC; AU Pr ; re Cr ; BA Pr ;details chapter 9.4). 
AU Pr 
5.  The changed setting is taken over by pressing the   key. 
6.  The balance returns to the weighing mode.  0.0 g 
8.5.2  Baud rate 
The baud rate necessary for the measuring values transfer can be set. 
In the following example the baud rate is set to 9600 baud. 
Setting of baud rate  Balance display 
1.   Keep the   key pressed until „Unit" is displayed. 
Unit 
2.    Press   key 
Pr 
3. Press    key, „baud“ appears. 
Baud 
4.  Confirm by pressing the   key 
4800 
5.  The baud rate can be changed using the   key  (1200, 
2400, 4800, 9600, 19200). 
9600 
6.  The changed setting is taken over by pressing the   key. 
7.  The balance returns to the weighing mode.  0.0 g 
440-BA-e-0941 
25