User manual
8.2  Leave menu structure 
Everywhere in the menu it is possible to leave the menu structure and thereby save 
or reject the changes made. 
After pressing the   key „Exit“ is displayed. 
A: Use the   (YES) key, to confirm. After that „store“ is displayed. If it shall be 
saved, press the   key repeatedly. 
If the menu shall be left without saving, 
press the   NO key. 
B: The   (NOT EXIT) key must be pressed, if the next menu item shall be 
reached. After having set all the individual adjustments, it can be saved. 
8.3  Dosing and Zero-tracking 
The Auto-Zero function is used to tare small variations in weight automatically. 
In the event that small quantities are removed or added to the material to be 
weighed, incorrect weighing results can be displayed due to the “stability 
compensation“ in the balance. (Example: Slowly draining fluids from a container on 
the balance). 
When apportioning involves small variations of weight, it is advisable to switch off this 
function. 
If Zero-Tracking however is switched off, the weighing display becomes more busy. 
Activate/deactivate Zero-Tracking  Balance display 
1.   Keep the   key pressed until „Unit" is displayed. 
Unit 
2.   Press the   key several times until „tr“ is displayed. 
tr 
3. Press the   key to activate the function. 
tr on 
4. By pressing once more the   key, the function is 
deactivated. 
tr off 
5.  The changed setting is taken over by pressing 
the 
 key. 
6.  The balance returns to the weighing mode.  0.0 g 