User manual
6.5  Connection of peripheral devices 
Before connecting or disconnecting of additional devices (printer, PC) to the data 
interface, always disconnect the balance from the power supply. 
With your balance, only use accessories and peripheral devices by KERN, as they 
are ideally tuned to your balance. 
6.6  Initial Commissioning 
In order to obtain exact results with the electronic balances, your balance must have 
reached the operating temperature (see warming up time chap. 1). During this 
warming up time the balance must be connected to the power supply (mains, 
accumulator or battery). 
 The accuracy of the balance depends on the local acceleration of gravity. 
Strictly observe hints in chapter Adjustment. 
6.7  Adjustment 
As the acceleration value due to gravity is not the same at every location on earth, 
each balance must be coordinated - in compliance with the underlying physical 
weighing principle - to the existing acceleration due to gravity at its place of location 
(only if the balance has not already been adjusted to the location in the factory). This 
adjustment process must be carried out for the first commissioning, after each 
change of location as well as in case of fluctuating environment temperature. To 
receive accurate measuring values it is also recommended to adjust the balance 
periodically in weighing operation. 
6.8  Adjustment 
The adjustment should be made with the recommended adjustment weight (see 
chap. 1 "Technical data"). Adjustment is also possible with the weights of other 
nominal values (see table 1), but not the optimum for measuring technique. 
Procedure when adjusting: 
Observe stable environmental conditions. A warming up time (see chapter 1) is 
required for stabilization. 
Switch on the balance using the   key. 
Press   key and keep it pressed, after the acoustic signal there appears in the 
display for short time „CAL“. After that the exact size appears flashing in the display 
(chapter.8.4) of the adjustment weight. 
Now set the adjusting weight in the centre of the weighing plate. 
Now press the   key. Short time later there appears „CAL F“, then the automatic 
return to the normal weighing mode. In the display there appears the value of the 
adjustment weight. 
An error during adjustment or the use of an incorrect adjusting weight will result in an 
error message „CAL E“. Repeat adjustment. 
Keep the adjustment close to the balance. Daily control of the weighing exactness is 
recommended for quality-relevant applications. 