Procedure when adjusting:
Attention:
In calibrated balances the adjustment function is switch locked.
To carry out calibrations, you will have to throw the unlock switch (See chpt. 11.1).
Observe stable environmental conditions. A warming up time of ca. 30 minutes is
required for stabilization. Ensure that there are no objects on the weighing plate.
Activate the calibrating function F3 in the menu (See chpt. 10)
Weight Basic price Sales price
F3 WIDE CAL.
Press -key, the following display will appear:
Weight Basic price Sales price
0,000 kg XXXXXX
(internal value)
Ensure that there are no objects on the weighing plate.
Operate menu activation key (10.1); stop check is carried out, then the weight
value of the calibrating weight will appear on the weight display
Weight Basic price Sales price
3.000
(maximum load)
Kg
(weighing unit)
XXXXXX
(internal value)
Carefully place adjusting weight in the centre of the weighing plate
Operate menu activation key (see chap. 10.1)
Weight Basic price Sales price
CAL. done
While the balance is counting down to zero, remove the adjusting weight.
The balance returns automatically into weighing mode. An error message will
appear on the display should a adjustment error occur or should the adjust-
ment weight be incorrect. Turn balance off, then restart it and repeat the ad-
justment process.
Notice: Again throw the verification switch for calibrated balances.