User Manual

PCB-BA-e-0914
5
KERN PCB 200-
2
PCB 250-
3
PCB 350-
3
PCB 400-
2
PCB 400-
1
Readability (d)
0,01 g 0,001 g 0,001 g 0,01 g 0,1 g
Weighing range (max) (d)
200 g 250 g 350 g 400 g 400 g
Taring range (subtractive)
200 g 250 g 350 g 400 g 400 g
Reproducibility
0,01 g
0,001 g
0,001 g 0,01 g 0,1 g
Linearity
± 0,02 g
±0,003 g
±0,003 g ±0,03 g ±0,2 g
Minimum unit weight at
piece counting
0,02 g 0,002 g
0,002 g
0,02 g 0,2 g
Warm-up time 30 min. 2 hours 2 hours 2 hours 10 min.
Reference quantities at
piece counting
5, 10, 20, 25, 50
Weighing Units
Details „Weighing units“ see chpt. 9.3
Recommended adjustment
weight, not added (class)
Details for „Selection of the
Adjustment weight“ in ch. 9.3
200g
(M1)
200g
(F1)
300g
(F1)
400g
(F2)
400g
(M 2)
Stabilization time (typical) 3 sec.
Operating temperature + 5° C .... + 35° C
Humidity of air max. 80 % (not condensing)
Housing (B x D x H) mm
163 x 245 x 79 without windscreen
163 x 245 x 123 with windscreen
Windscreen -
Ø 90,
53 high
Ø 90,
53 high
- -
Weighing plate mm Ø 105 Ø 81 Ø 81 Ø 105
130 x 130
Weight kg (net) 1,1 1,4
Electric Supply 220V-240V AC, 50 Hz / 9 V, 300 mA
Battery operation 9 V compound battery (optional)
Rechargeable battery (optional)
Operating time 24 h with display background
illumination
Operating time 48 h without display background
illumination/loading time 8 h
Interface RS 232
Underfloor weighing standard