User Manual
PCB-BA-e-0914
4
1 Technical Data
KERN PCB 40-3 PCB 60-3 PCB 100-3 PCB 160-3
Readability (d) 0,001 g 0,001 g 0,001 g 0,001 g
Weighing range (max) (d) 40 g 60 g 100 g 160 g
Taring range (subtractive) 40 g 60 g 100 g 160 g
Reproducibility 0,001 g 0,001 g 0,001 g 0,001 g
Linearity ±0,003 g ±0,003 g ±0,003 g ±0,003 g
Minimum unit weight at
piece counting
0,002 g 0,002 g 0,002 g 0,002 g
Warm-up time 2 hours
Reference quantities at
piece counting
5, 10, 20, 25, 50
Weighing Units
Details „Weighing units“ see chpt. 9.3
Recommended adjustment
weight, not added (class)
Details for „Selection of the
Adjustment weight“ in ch. 9.3
40g (F1) 50g (F2) 100 g (F1) 150 g (F1)
Stabilization time (typical) 3 sec.
Operating temperature + 5° C .... + 35° C
Humidity of air max. 80 % (not condensing)
Housing (B x D x H) mm
163 x 245 x 79 without windscreen
163 x 245 x 123 with windscreen
Windscreen Ø 90, 53 high
Weighing plate mm Ø 81
Weight kg (net) 1,1
Electric Supply 220V-240V AC, 50 Hz / 9 V, 300 mA
Battery operation 9 V compound battery (optional)
Rechargeable battery (optional)
Operating time 24 h with display background
illumination
Operating time 48 h without display background
illumination/loading time 8 h
Interface RS 232
Underfloor weighing standard