Manual

9 Technical examination
Technical examination by expert staff is required every three
months.
The examination should include the following:
Condition of connection to earth for balance display unit, including:
o Fastening of display housing and electrical resistance - max. 100Ω,
o Fit a flexible steel tube for the protection of the conductor leading to the
platform balance and the display unit (regular balances with “cabled”
display).
The quality of the connection between the platform and the display unit (flexible
steel tube connecting both subassemblies).
Damage and loss of a connection is inacceptable.
The electrical connection between the housing and the platform the maximum
resistance between the weighing pan (put on) should be less than100Ω.
Stability and levelling of the balance.
Doc.No.OEX-BA-e-1411 15