KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-Mail: info@kern-sohn.com Tel: +49-[0]7433- 9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com Operating instructions Personal floor scale with BMI function KERN MPC_M / MPE_HM / MPE_PM Version 1.
GB KERN MPC 250K100M KERN MPE 250K100HM KERN MPE 250K100PM Version 1.4 10/2014 Operating instructions Personal floor scale with BMI function Contents 1 Technical data ............................................................................................ 4 2 Declaration of conformity .......................................................................... 6 2.1 Explanation of the graphic symbols ........................................................................ 6 3 Appliance overview .
10 Unpacking, Setup and Commissioning .................................................. 23 10.1 10.2 10.3 Installation Site, Location of Use ...........................................................................23 Unpacking .............................................................................................................23 Scope of delivery ..................................................................................................24 10.3.1 10.3.2 10.3.3 MPC models ..............
1 Technical data KERN MPE 250K100HM MPE 250K100PM Display 6-digit Weighing range (max) 250 kg Minimum load (Min) 2 kg Verification value (e) 100 g Reproducibility 0.1 kg Linearity ± 0.1 kg Display LCD with 25mm high digits > 200 kg (M1) Recommended adjustment weight, (Class) Stabilization time (typical) 2 sec. Warm-up time 10 min Operating temperature 0° C …. + 40° C Humidity of air max.
KERN MPC 250K100M Display 6-digit Weighing range (max) 250 kg Minimum load (Min) 2 kg Verification value (e) 100 g Reproducibility 0.1 kg Linearity ± 0.1 kg Display LCD with 25mm high digits > 200 kg (M1) Recommended adjustment weight, (Class) Stabilization time (typical) 2 sec. Warm-up time 10 min Operating temperature 0° C …. + 40° C Humidity of air max.
2 Declaration of conformity Declaration of conformity: see separate document showing serial number of device CE marking: 93/42/EEC 0297 year 0103 2.1 M 2009 / 23 / EG Non-automatic Weighing Instruments Directive Explanation of the graphic symbols This EC verification mark indicates that these scales are in conformity with EU Directive 2009/23/EG for Non-Automatic Weighing Instruments. Weighing instruments bearing this mark are approved for medical purposes within the European Union.
Identification of manufacturer of medical product including address Kern & Sohn GmbH D–72336 Baligen,Germany www.kern-sohn.com “Electro-medical appliance“ with attachment for type B Device protection category II Dispose of old appliances separately from your household waste! Instead, take them to communal collection points.
3 Appliance overview 3.1 MPC models: 1 1. Weighing platform (anti-slip surface) Underside 2. Rubber feet (height adjustable) 3. Bubble level 4 5 6 4. Terminal power supply unit 5. Connecting cable ”display unit - platform" 6.
3.2 MPE models: 1. Height measuring stick (MPE-HM models only) 2. Display Unit 3. Weighing platform (anti-slip surface) 4. Rubber feet (height adjustable) Underside 5.
Secondary display at rear Display unit at rear 5 Rechargeable battery compartment 6 Mains connection 7 RS 232 C 10 MPC_M / MPE_HM / MPE_PM-BA-e-1414
4 Keyboard overview 4.1 MPC models Key Description Function ON/OFF-switch Turn on/off HOLD button Hold function/Calculation of a stable weight value BMI key Calculation of the Body Mass Index In menu: • Confirm selection For numeric entry: • Confirm numerical value Function key In menu: • Call up menu • How to select menu items For numeric entry: • Zeroing key Weighing scale will be reset to „0.
4.2 MPE models Key Description Function ON/OFF-switch Turn on/off HOLD button Hold function/Calculation of a stable weight value BMI key Calculation of the Body Mass Index PRINT button Data transfer via interface In menu: • Confirm selection For numeric entry: • Confirm numerical value (MPE models only) 12 Function key In menu: • Call up menu • How to select menu items For numeric entry: • Increase numerical value Zeroing key Weighing scale will be reset to „0.
5 Overview of display Display Description Description Stability display Scales are in a steady state Zeroing display Should the balance not display exactly zero despite empty weighing plate, press the button. Your balance will be set to zero after a short standby time.
6 Basic Information (General) Weighing instruments have to be verified for the purposes stated below in accordance with Directive 2009/23/EC. Article 1, paragraph 4. “Determination of mass in the practice of medicine that is, weighing patients for reasons of medical supervision during medical surveillance, examination and treatment.” 6.1 Specific function Indication • • Determining the body weight in the medical practice area.
6.3 Improper Use Do not use these scales for dynamic weighing processes. Do not leave permanent load on the weighing pan. This may damage the measuring system. Impacts and overloading exceeding the stated maximum load (max) of the weighing plate, minus a possibly existing tare load, must be strictly avoided. This could cause damage to the balance. Never operate balance in explosive environment. The serial version is not explosion protected.
7 Basic Safety Precautions 7.1 Pay attention to the instructions in the Operation Manual Carefully read this operation manual before setup and commissioning, even if you are already familiar with KERN balances. All language versions contain a non-binding translation. The original German is binding. 7.2 Personnel training The medical staff must apply and follow the operating instructions for proper use and care of the product. 7.
8 Electromagnetic compatibility (EMC) 8.1 General hints The installation and use of this electrical medical device requires special precautionary measures as outlined in the EMC information below. This device complies with the limits set for medical electrical devices of group 1, class B (as per EN 60601-1-2). Electromagnetic compatibility (EMC) describes a device’s ability to perform reliably within an electromagnetic environment without causing inadmissible electromagnetic interference at the same time.
8.2 Electromagnetic interferences Guidelines and manufacturer’s declaration – electromagnetic interferences The floor scales MPC/MPE is designed for use in an electromagnetic environment that meets the requirements stated below. The customer or user of the medical electrical device must ensure that operation takes place in such an environment.
8.3 Electromagnetic noise immunity Guidelines and manufacturer’s declaration - electromagnetic noise immunity The floor scales MPC/MPE is designed for use in an electromagnetic environment that meets the requirements stated below. The customer or user of the medical electrical device must ensure that operation takes place in such an environment.
Guidelines and manufacturer’s declaration - electromagnetic noise immunity The floor scales MPC/MPE is designed for use in an electromagnetic environment that meets the requirements stated below. The customer or user of the medical electrical device must ensure that operation takes place in such an environment. Noise immunity tests IEC 60601 test level Conformity Electromagnetic environment guideline Conducted HF disturbance variables 3 150 kHz to 80 MHz 3V 3 80 MHz to 2.
8.3.1 Crucial features of performance Note: The floor scales MPC/MPE does not have any crucial features of performance as per IEC 60601-1. The system may be subject to interference by other devices even if these devices conform to current emission requirements as per CISPR. 8.
9 Transport and storage 9.1 Testing upon acceptance When receiving the appliance, please check packaging immediately, and the appliance itself when unpacking for possible visible damage. 9.2 Packaging / return transport Keep all parts of the original packaging for a possibly required return. Only use original packaging for returning. Prior to dispatch disconnect all cables and remove loose/mobile parts. Reattach possibly supplied transport securing devices.
10 Unpacking, Setup and Commissioning 10.1 Installation Site, Location of Use The balances are designed in a way that reliable weighing results are achieved in common conditions of use. You will work accurately and fast, if you select the right location for your balance.
10.3 Scope of delivery Serial accessories: 10.3.1 MPC models • Balance • Power pack unit (EN 60601-1 attestation of conformity) • • Operating instructions Wall bracket • Protective working cover 10.3.2 MPE-HM models • Balance • Power pack unit (EN 60601-1 attestation of conformity) • Operating instructions • • Height measuring stick Protective working cover 10.3.
10.4.2 MPE models (Example shown for MPE-HM model) Scope of delivery: Fig.
Refitting Procedure: Mount tripod (1) on tripod holder (2) at weighing platform Ensure that the cable (3) does not get clamped! Fasten tripod, using 4 screws 10.5 Mains connection Power is supplied by the external power unit which also serves to isolate the mains supply from the scale. The stated voltage value must be the same as the local voltage. Always use genuine approved KERN power pack units as per EN 60601-1 directive.
10.6 Battery operation is possible by obtaining an optional battery power pack. MPC 1 MPE 1 Open the battery compartment cover (1) at the base of the display unit and insert the rechargeable battery. Charge the battery for at least 12 hours before initial use. The appearance of the symbol in the weight display indicates that the battery is almost exhausted. The weighing scale will remain ready for operation for a few more minutes before switching off in order to save battery (s. chap. 11.6 Auto off).
10.7 Battery operation As an alternative to rechargeable battery operation, the balance may also be operated with 6x AA batteries. Open battery compartment cover (1) at the lower side of the display unit and insert batteries according to the example below. Lock the battery cover again. If the batteries are empty, in the balance display appears the symbol . Change batteries. To save battery power, the balance switches off automatically (see chap.11.6 Auto off). Capacity of batteries exhausted.
Insert batteries Version A: Remove the battery compartment cover and detach the adhesive tape carefully. Ensure that the cables are not destroyed. Insert contact sheet and sink the battery compartment connection into the provided opening (see fig. below) Contact sheet Plug for rechargeable battery connection Opening for rechargeable battery connection (Should the rechargeable battery be used later, take care that the plug does not disappear in the housing) Clamp the contact sheet acc.
Version B: Remove battery compartment cover Connect battery holder to housing contact acc. to illustration Insert battery holder Insert batteries into battery compartment and lock with battery compartment cover. 10.8 Initial Commissioning In order to obtain exact results with the electronic balances, your balance must have reached the operating temperature (see warming up time chap. 1).
11 Operation 11.1 Weighing Start balance by pressing The balance will carry out a self-test The scales are ready for operation as soon as the weight display for “0.0 kg“ has appeared. • However, you can reset the weighing scale to zero by pressing the key. Have person stand in the centre of the scales. Wait until the standstill display „STABLE“ appears, then read the weighing result. • If the person is heavier than the weighing range, "OL" (=overload) will appear in the display. 11.
11.2.1 Subsequent tare weight The balance can be tared several times successively. 11.3 HOLD function The balance has an integrated standstill function (mean value calculation). With this function it is possible to weigh people accurately even if they do not stand still on the weighing plate. Start balance by pressing . Wait for stability display „STABLE to appear. Press (MPE) (MPC) in the display „---------“ will appear and the „HOLD“ symbol appears.
11.5 Calculation of the Body Mass Index You need to know a person’s body height before you can calculate the BMI for that person. It should either be known or can be determined directly with the MPE-HM model. 11.5.1 Measuring body height (MPE-HM only) Push measuring stick upwards and set the stopper horizontally. Push measuring rod carefully down until the headpiece touches the person’s head. (It is recommended to take measurements with shoes removed).
Read body height on measuring stick. 11.5.2 Calculating Body Mass Index Start balance by pressing Wait for stability display „STABLE to appear. Press (MPE) or (MPC). The most recently entered body height will be shown with the enable digit flashing. The „BMI“ symbol lights up. To enter body height, press the and Confirm your entered value by pressing key. (MPE) or (MPC). BMI „0.0“ is displayed Allow the person to step onto the centre of the weighing platform.
• • Reliable calculation of BMI is restricted to a body height of 100 cm to 200 cm and a weight of >10 kg. If weighing has to take place under unsteady conditions, you can be stabilise the display by applying the Hold function. 11.5.3 Classification of BMI values Weight classification for adults over 18 years of age using the BMI in accordance with WHO, 2000 EK IV and WHO 2004. Categorie BMI ( kg/m2) Risk of diseases associated with overweight Underweight < 18.5 low Normal weight 18.5 – 24.
11.6 Automatic switch-off function „AUTO OFF“ The weighing scale will switch off automatically after the allotted time as long as neither the display unit nor the weighing plate is operated. • Menu settings: [F1 oFF] [oFF 0/3/5/15/30] (see chap. 12) Start balance by pressing . Wait for stability display „STABLE to appear.
11.7 Display background illumination • Menu settings: [F4 bk] [bL on / bL oFF / bL AU] see chap. (12) • Menu settings:MPC: [F4 or F2 bk] [bL on / bL oFF / bL AU] see chap. (12) Start balance by pressing . Wait for stability display „STABLE to appear. Press , [F1 oFF] is displayed Press repeatedly until [F4 bk] is shown.
12 The menu Access to service menu „tCH“is locked in verified weighing scales. To disable the access lock, destroy the seal and actuate the adjustment switch. For position of adjustment switch, see chap. 16. Attention: After destruction of the seal the weighing system must be re-verified by an authorised agency and a new verification wire/seal mark fitted before it can be reused for applications subject to verification. 12.
12.2 Menu overview MPC models 12.2.1 Models MPC without RS232 interface Function F1 oFF Automatic cutout Auto Off F2 bk Background illumination of display F3 Str Subsequent tare value locked in devices with type approval certificate.
12.2.2 Models with RS232 interface Function Settings Description Automatic cutout Auto Off Background illumination of display Background illumination of display oFF 0* Automatic shutdown off oFF 3 Automatic shutdown after 3 min. oFF 5 Automatic shutdown after 5 min. oFF 15 Automatic shutdown after 15 min. oFF 30 Automatic shutdown after 30 min.
3. Data output format (Setting P Prt, P Auto, P Cont only!) The currently set data output format will be displayed after the Baud rate has been confirmed. Select desired format by and confirm with Data output format, see chap. 13.3 Cont 1 Default Cont 2 Not documented Cont 3 Not documented Only when set P Cont only at setting P Prt, P Prt 0-3 . Sd0 – on/off Continuous data output, selectable “Sending 0” yes / no 4.
12.3 Menu overview MPE models Menu block Main Menu Menu item Submenu Available settings / explanation oFF 0* Automatic shutdown off oFF 3 Automatic cutout oFF 5 Auto Off Automatic shutdown after 3 min Automatic shutdown after 5 min oFF 15 Automatic shutdown after 15 min oFF 30 Automatic shutdown after 30 min oFF Not documented Prt Pr ACC 1. RS-232 mode Interface parameter Select desired mode by P Prt P Cont Series ASK , then confirm with .
3. Data output format (P Prt, P Auto, P Cont settings only) the currently set data output format will be shown after the baud rate was confirmed. Select desired format by Only when set P Cont Only when set P Prt, P confirm with and . Prt 0-3 Data output format, see chap. 13. Cont 1 Default Cont 2 Not documented Cont 3 Not documented Sd0 – on/off Cont. data outputselectable „send 0“, yes / no 4.
Pin Service menu Password entry: press subsequently. , , Operate adjustment switch; for position see chap.18 15* Display speed 30 60 Not documented 7.5 Adjustment, see chap.
13 Data output RS 232 You can print weighing data automatically via the RS 232 interface or manually by pressing mit in the menu. (MPE) bzw. (MPC) via the interface according to the setting This data exchange is asynchronous using ASCII - Code. The following conditions must be met to provide successful communication between the weighing balance and the printer. • Use a suitable cable to connect the weighing balance to the interface of the printer.
13.3 Printer operation Printout examples: Prt 0/2 60.0kg 1/3 60.0kg 170.0cm 20.7BMI Remote control instructions: S Stable value positive Stable value negative ST,GS ST,NT- 10.0kg 20,5kg W Instable value positive Instable value negative US,NT 33.0kg US,NT- 20,5kg Taring ST,GS 0.
14 Error messages Display Description Zero range exceeded (on start-up or when pressing the key) • Load on weighing pan • Excess load, during zero setting of weighing scale • Incorrect adjusting process • Fault on load cell Value outside the A/D converter range • Damaged weighing cell • Damaged electronics Should other error messages occur, switch balance off and then on again. If the error message remains inform manufacturer.
15 Service, maintenance, disposal 15.1 Cleaning Before any maintenance, cleaning and repair work disconnect the appliance from the operating voltage. 15.2 Cleaning / disinfecting Clean weighing platform (such as seat) as well as casing with household detergents or commercially available disinfectants. Please follow manufacturer’s instructions. Do not use abrasive or aggressive cleaners such as spirits or alcohol or similar as they might damage the high-quality surface.
16 Instant help In case of an error in the program process, briefly turn off the balance and disconnect from power supply. The weighing process must then be restarted from the beginning. Fault Possible cause The displayed weight does • The balance is not switched on. not glow. • The mains supply connection has been interrupted (mains cable not plugged in/faulty). • Power supply interrupted.
17 Verification General introduction: According to EU directive 2009/23/EC balances must be officially verified if they are used as follows (legally controlled area): a) For commercial transactions if the price of goods is determined by weighing. b) For the production of medicines in pharmacies as well as for analyses in the medical and pharmaceutical laboratory. c) For official purposes d) For manufacturing final packages In cases of doubt, please contact your local trade in standard.
Position adjustment switch and seals: 1 2 3 1 KERN MPE 1 1 2 3 KERN MPC 1. 2. 3.
17.1 Verification validity period (current status in G) Personal scales (including chair and wheelchair scales) in hospitals 4 year Personal scales, when not located in hospitals (for example, doctor's offices and nursing homes) unlimited Baby weighing scales and mechanical birth weight scales 4 year Bed scales 2 year Scales in dialysis stations unlimited Rehab clinics and health authorities are treated as hospitals.
18 Adjustment As the acceleration value due to gravity is not the same at every location on earth, each display unit with connected weighing plate must be coordinated - in compliance with the underlying physical weighing principle - to the existing acceleration due to gravity at its place of location (only if the weighing system has not already been adjusted to the location in the factory).
KERN MPE Press , and one after the other and [P1 SPd] will appear KERN MPC Press appear , Press and one after the other and [P1 SPd] will , „P2 CAL“ will be displayed Operate adjustment switch; for position see chap.16 Press (MPE) or Press repeatedly until „CAL“ will be displayed. To confirm, press (MPC) and [dESC] will appear (MPE) or (MPC) and [UnloAd] will appear Ensure that there are no objects on the weighing pan.
Place adjustment weight in the centre of the weighing pan Wait until stability display „STABLE“ appears Confirm by pressing (MPE) or (MPC), [PASS] will be displayed. The balance carries out a selftest, after that [Err19] will be displayed and a signal will sound. Switch off the balance Take away adjustment weight Turn on balance again, after the selftest the balance changes into the weighing mode. Adjustment has now been completed successfully.