Owner's manual

16 Verification
Verified scales bear a verification mark or one or more seals affixed by the Bureau of
Standards or the manufacturer on or inside the housing which will self-destroy on
removal. This makes it impossible to verify scales without damaging the seals.
16.1 Adjustment
Observe stable environmental conditions. A warming up time (see chapter 1) is
required for stabilization.
In calibrated balances the adjustment function is switch locked.
In order to carry out adjustments, the switch must be turned to
adjustment position (centre position). (see chapter 15.2).
Display
Operation
a
z000
��
Start balance by pressing
OFF
ON
SETUP ��
UNKT ��
Press for approx. 3 sec,
in the display appears „SETUP“, followed by „UNIT“
CAL ��
K
B
Press
TARE
repeatedly until “CAL iB“ appears
CAL U ��
Press
HOL D
, „CAL U“ appears
Press
TARE
, the appeared triangle must be located in the
upper left part of the display.
If not, press again
TARE
.
MCB_MPT-BA-e-1418 45