KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen email: info@kern-sohn.com Phone: +49-[0]7433- 9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com Operating instructions Medical chair scales Medical personal scales KERN MCB, MPT Version 1.
GB KERN MCB, MPT Version 1.8 03/2014 Operating instructions chair scales, personal scales Table of Contents 1 Technical Data ............................................................................................... 4 2 Declaration of conformity ............................................................................. 5 2.1 Explanation of the graphic symbols ....................................................................................... 5 3 Basic Information (General) .............
11 Menu overview ............................................................................................. 32 12 Operation ...................................................................................................... 33 12.1 12.2 12.3 12.4 12.4.1 12.5 12.5.1 12.6 12.6.1 Weighing ................................................................................................................................. 33 Taring ..........................................................................
1 Technical Data KERN MCB 300K100M MPT 300K100M Readability (d) 100 g Weighing range (max) 300 kg Minimum weight (min) 2 kg Verification value (e) 100g Verification class III Recommended adjustment weight, (Class) 300 kg (M1) Weighing Units kg Stabilization time (typical) 2-3 sec.
2 Declaration of conformity Declaration of conformity: See separate document showing serial number of device CE marking: 93/42/EEC 0297 year M 0103 2.1 2009/23/EC Non-automatic Weighing Instruments Directive Explanation of the graphic symbols This EEC verification mark indicates that these scales are in conformity with EEC Directive 2009/23/EC for Non-Automatic Weighing Instruments. Weighing instruments bearing this mark are approved for medical purposes within the European Union.
Please note operating instructions Please note operating instructions “Electro-medical appliance“ with attachment for type B Device protection category II Dispose of old appliances separately from your household waste !!! Instead, take them to communal collection points.
3 Basic Information (General) Weighing instruments have to be verified for the purposes stated below in accordance with Directive 2009/23/EC. Article 1, paragraph 4. “Determination of mass in the practice of medicine that is, weighing patients for reasons of medical supervision during medical surveillance, examination and treatment.” 3.1 Specific function Indication Determining the body weight in the medical practice area.
When the scale doesn’t connect with the transmission cable, please do not touch the transmission port to prevent ESD interference occurs. The chair scales may not be used for the transport of people! As long as the patient is remaining on the chair scales, the wheel brakes must be locked without fail. Do not step onto the foot rests when stepping or leaving the chair scales! 3.3 Improper Use Do not use these scales for dynamic weighing processes.
3.4 Warranty Warranty claims shall be voided in case Our conditions in the operation manual are ignored The appliance is used outside the described uses Modification or opening of appliances mechanical damage and damage caused by media, liquids, Natural wear and tear The appliance is improperly set up or incorrectly electrically connected The measuring system is overloaded Dropping of scales 3.
4 Basic Safety Precautions 4.1 Pay attention to the instructions in the Operation Manual Carefully read this operation manual before setup and commissioning, even if you are already familiar with KERN balances. Versions in other languages are nonbinding translations. The only binding version is the original document in German. 4.2 Personnel training The medical staff must apply and follow the operating instructions for proper use and care of the product. 4.
5 EMC guidance and manufacturer's declaration Guidance and manufacturer’s declaration-electromagnetic emissions The MCB 300K100M, MPT 300K100M is intended for use in the electromagnetic environment specified below. The customer or the user of the MCB 300K100M, MPT 300K100M should assure that it is used in such an environment.
Guidance and manufacturer’s declaration-electromagnetic immunity The MCB 300K100M, MPT 300K100M is intended for use in the electromagnetic environment specified below. The customer or the user of the MCB 300K100M, MPT 300K100M should assure that it is used in such an environment.
Guidance and manufacturer’s declaration-electromagnetic immunity The MCB 300K100M, MPT 300K100M is intended for use in the electromagnetic environment specified below. The customer or the user of the MCB 300K100M, MPT 300K100M should assure that is used in such and environment.
Recommended separation distance between portable and mobile RF communications equipment and the MCB 300K100M, MPT 300K100M The MCB 300K100M, MPT 300K100M is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
6 Appliance overview Chair scales MCB 1 2 3 6 4 5 1. 2. 3. 4. 5. 6.
Personal scales MPT 1 2 3 4 2 5 1. 2. 3. 4. 5.
7 Overview of displays Display 0 Designation Description Stability display Scales are in a steady state Zeroing display Should the balance not display exactly zero despite empty scale pan, press the button. Your balance will be set to zero after a short standby time.
8 Keyboard overview Button ON OFF PRINT BMI HOLD PRETARE TARE CLEAR ENTER Designation Function ON/OFFswitch Turn on/off PRINT button Data transfer via interface BMI button Calculation of the Body Mass Index HOLD button Hold function/Calculation of a stable weight value Zero setting key Balance will be reset to 0.0 kg. Possible up to max.
9 Transportation & Storage 9.1 Testing upon acceptance When receiving the appliance, please check packaging immediately, and the appliance itself when unpacking for possible visible damage. 9.2 Packaging / return transport Keep all parts of the original packaging for a possibly required return. Only use original packaging for returning. Prior to dispatch disconnect all cables and remove loose/mobile parts. Reattach possibly supplied transport securing devices.
10 Unpacking, Setup and Commissioning 10.1 Place of installation / place of operation The balances are designed in a way that reliable weighing results are achieved in common conditions of use. You will work accurately and fast, if you select the right location for your balance.
10.4 Installing the chair scales In delivery state the balance is adjusted in a manner that when placed on an even surface the air bubble of the water balance is in the prescribed circle. To check place scales on an even surface. Check if the air bubble of the level is within the prescribed circle. If the air bubble in the level is not within the prescribed circle, the wheel height has to be adjusted, see chap. 9.4.1. Check levelling regularly.
10.4.1 Levelliing • The wheel height has to be adjusted for levelling. Levelling is restricted to specialist staff possessing well acquainted • with the workings of weighing scales. Place scales on an even surface. Arrest the brake Turn the wheel (1) clockwise until the air bubble of the water balance is in the prescribed circle Screw counternut (2) upwards till to the stop and it with a suitable tool (e.g. pliers).
10.5 Scope of delivery, personal scales Balance Operating instructions Mains adapter Wall bracket 4 rubber feet 10.6 Installing the personal scales Place scales on an even surface Level balance with foot screws until the air bubble of the water balance is in the prescribed circle. 10.
10.8 Attaching an optional stand Secure circular plate with screws on the aluminium section Secure wall fixing device with screws at the top on the aluminium section Remove the two lateral rubber plugs from the display unit. Secure the display unit with the two rotary knobs to the fixing device. Position display unit with the rotary knobs. Secure cable with cable clips 10.
Ensure that the batteries are inserted in the correct direction Insert the battery holder with the inserted batteries in the display unit Close the battery cover If the batteries are run down, "LO" appears in the display. ON To turn off scales, press the OFF button and immediately change the batteries. If the balance is not used for a longer time, take out the batteries and store them separately. Leaking battery liquid could damage the balance.
10.
10.10.1 Battery operation On models where the back of the display unit is not directly accessible, remove the two black rotary knobs from both sides of the display unit in order to open the battery compartment and remove the display unit from the holder. Lift-off the battery cover on the lower side of the balance Carefully take out the battery holder (1) Insert 6 batteries (AA).
Insert battery holder with the inserted batteries into the display unit Ensure that the cables are not squeezed Close the battery cover If the batteries are run down, "LO" appears in the display. To turn ON off, press the OFF button and immediately change the batteries. If the balance is not used for a longer time, take out the batteries and store them separately. Leaking battery liquid could damage the balance.
10.10.2 Rechargeable battery operation (optional) When an optional rechargeable battery is used, proceed as follows: On models where the back of the display unit is not directly accessible, remove the two black rotary knobs from both sides of the display unit in order to open the battery compartment and remove the display unit from the holder.
Carefully insert the rechargeable battery block and insert plug into connection CN 3 Ensure that the cables are not squeezed Close the battery cover If the rechargeable battery is exhausted, „LO“ is displayed. The rechargeable battery is loaded via the provided plug-in power supply unit (loading time 14 h for a complete loading). If the balance is not used for a longer time, take out the rechargeable battery and store it separately. Leaking liquid could damage the balance.
10.11 Connecting the power supply Power is supplied via the external mains adapter which also serves to isolate the scales from the mains. The stated voltage value must be the same as the local voltage. Only approved genuine KERN power supply units may be used in compliance with Directive EN 60601-1. 10.12 Initial Commissioning In order to obtain exact results with the electronic balances, your balance must have reached the operating temperature (see warming up time chap. 1).
11 Menu overview a Start balance by pressing z000 �� SETUP Press ON OFF for 3 sec., “SETUP“ will be displayed. Select parameter by using described. TARE HOLD ( ) and Confirm the selected parameter by pressing TARE ( ), as ( ). 0.0 kg AOFF Auto off: 120 sec / 180 sec / 240 sec / 300 sec / OFF bUrr Audio signal: ON/OFF End Exit menu HOLD If End is selected, finish set up by pressing the 32 button.
12 Operation 12.1 Weighing P-1.32 0 0.0 ON OFF Start balance by pressing The balance carry out a segment test, then the program version is displayed. The scales are ready for operation as soon as the weight display for “0.0kg“ has appeared. Chair scales Personal scales The time. button can be used to set the balance to zero at any Have person sit in the centre of the scales. Fold down the foot rests and place the patient’s feet on the respective foot rest.
The balance is able to only store one taring value. When the balance is unloaded the saved taring value is displayed with negative sign. To delete the stored tare value, release scales and press TARE . 12.3 Hold function (Standstill function) The balance has an integrated standstill function (mean value calculation). This allows correct weighing determination of a person although the latter is not keeping still on the scales. 0 0.0 HOLD 62.
12.4 Calculation of the Body Mass Index 0.0 0 0 15.0 ON OFF Start balance by pressing Wait for the rest position display . Have person sit or stand in the centre of the seating surface or weighing platform. Wait for the rest position display . BMI 165.0 Press The first decimal place of the most recently entered body height is flashing. It should be noted that a reliable calculation of the BMI is only possible with a body height of between 100 cm and 250 cm and a weight >10kg.
12.4.1 Classification of BMI values Weight classification for adults over 18 years of age using the BMI in accordance with WHO, 2000 EK IV and WHO 2004. 2 Categorie BMI ( kg/m ) Risk of diseases associated with overweight Underweight < 18,5 low Normal weight 18.5 – 24.9 Average Overweight > 25.0 Pre-adipose 25.0 – 29.9 A bit high Adipose degree I 30.0 – 34.9 High Adipose degree II 35.0 – 39.
12.5 PRE-TARE function If the tare weight is known (rubber mat, clothing…..) . you may inter the value manually. 0 0.0 ON Start balance by pressing OFF Wait for the rest position display . PRETARE 000.0 Press briefly. A flashing display appears. A small arrow will be shown on the display under “PRE-TARE” and under “NET” as long as the PRE-Tare function is enabled. The most recently applied value or “000.0“ will appear. Go to the numeric keypad and enter a new value.
12.5.1 PRE-TARE function with 5 memories This gives the option of storing 5 different tare values that can be retrieved when required. Saving PPE-TARE values 0 0 0.0 10.0 Weighing platform is unloaded and indicates “0.0 kg“. Place the weight to be saved on the weighing platform and wait until the weight display is steady. Keep pressing the button until “ni“ (M) appears. Briefly press the numeric key ( 1..5 ) that you wish to use for saving the value.
Retrieving PPETARE values from the memory 0 0.0 Weighing platform is unloaded and indicates “0.0 kg“. PRETARE Press and hold the button until “ni“ appears and starts flashing on the display. 10.0 Briefly press the numeric key ( 1..5 ) under which number the Pre-Tare value was saved. The saved weight value will be flashing. In addition there will be a flashing triangle under “PRETARE“ and “NET“. Import value with ENTER . 0 10.0 92.5 Value will be shown with a negative sign.
PREPrintout from PRETARE and hold until “ni“ (M) appears on the display. TARE memory Press To issue the values saved to the 5 memories, briefly press PRINT the button. Example: M1 0.0 kg M2 7.0 kg M3 10.0 kg M4 30.0 kg M5 50.0 kg 12.6 Print function This action requires the RS232 interface cable available as an accessory which has to be connected via the round plug to the rear end of the terminal.
12.6.1 Interface parameters for RS232 The interface parameters of the scales have to be set at the connected appliance. It is not possible to change the parameters inside the scales.
13 Service, maintenance, disposal 13.1 Cleaning Disconnect the unit from the mains power supply prior to cleaning. 13.2 Cleaning/disinfection Clean the scale platform (e.g. the detachable plate) only with household detergent or commercially available disinfectant. Follow the manufacturer’s recommendations. Do not use abrasive or aggressive cleaning agents like spirit, gasoline or similar as they may damage the high quality surface.
14 Error messages Display Description LQ ERRL ERR ERRP Battery low Please replace the batteries or connect scales to the power supply, using an adaptor. Underload Weight on weighing platform is too low. Please increase weight. If the error message remains inform manufacturer. Overload Weight on weighing platform is too heavy. Program error Please contact your supplier.
15 Instant help In case of a fault in the program sequence, the balance should be shortly switched off. The weighing process must then be restarted from the beginning. Failure: Possible cause: The displayed weight does not glow. The displayed weight is permanently changing The weighing result is obviously incorrect The balance is not switched on. The mains supply connection has been interrupted (mains cable not plugged in/faulty).
16 Verification Verified scales bear a verification mark or one or more seals affixed by the Bureau of Standards or the manufacturer on or inside the housing which will self-destroy on removal. This makes it impossible to verify scales without damaging the seals. 16.1 Adjustment Observe stable environmental conditions. A warming up time (see chapter 1) is required for stabilization. In calibrated balances the adjustment function is switch locked.
CAL U �� Press HOLD repeatedly until “CAL 0“ appears. CAL O �� 30770 TARE Press , the display will show a numeric value. ENTER Then press , CAL O �� “CAL 0“ appears. HOLD CAL 5 �� Press , “CAL 5“ appears. 200.0 TARE Press Enter the required size of the adjustment weight (See Section 1, "Technical Data"): HOLD For this select the digit to be changed with 250.0 numeric value using Acknowledge by 82077 . ENTER .
16.2 Adjustment controls and seals After a verification the balance is sealed at the indicated positions. Verification of the balance is invalid without the "seal". Position of the official seals: 1. Rear MCB_MPT-BA-e-1418 2.
Position of the adjustment switch: 1 Position of the adjustment switch 2 3 State 1. to the left Not documented 2. concentric Adjustment position – adjustment possible 3. to the right Verification position - Adjustment locked 16.3 Checking the balance verification settings For the adjustment function, the balance must be switched over to service mode. To achieve the effect, turn the adjustment switch to adjustment position. In the service mode all parameters of the balance can be modified.
16.4 Navigation in the menu With the balance switched on, keep the button depressed for approx. 3 sec until "SETUP" is displayed followed by "9rAd". Press the displayed. Press the displayed. Press the the 16.4.1 TARE TARE button as often as necessary until the required function is HOLD button to confirm the selected function. The first parameter will be HOLD button to select the required parameter and confirm by pressing button.
HOLD and Selection available via TARE button * * * * * * * * * * * * default setting 50 MCB_MPT-BA-e-1418
Description: Weighing unit: kg partition steps, weighing range (max.) and readout (d) Multi-range balance/ single-range balance selection Single-range balance Multi-range balance Filter: fast/ normal/ slow Auto Zero Tracking: 0.25d / 0.5d / 1d / 3d / OFF Stabilisation range: 0.25d / 0.5d / 1d / 3d / OFF Zero range: 2% / 100% Overload range: 9d / 2% Adjustment Auto off: 120 sec. / 180 sec. / 240 sec. / 300 sec.
16.5 Verification validity period (current status in G) Personal weighing scales in hospitals 4 years Personal weighing scales if placed outside hospitals without time Baby and mechanical scales Infant scales 4 years Bed scales 2 years Wheelchair scales 2 years The hospitals also include rehabilitation clinics and health centres (4-year validity of verification). The hospitals do not include dialysis centres, care homes and consultation rooms (verification validity without time limit).