11 Troubleshooting
The scale should be switched off for a short time following an interruption in the pro-
gram sequence and disconnected from mains supply. It is then necessary to repeat
the weighing process from the beginning.
Help:
Interruption Possible cause
Weight display is not illumi-
nated.
The scale is not switched on.
Batteries are incorrectly inserted or discharged
No batteries.
Weight display changes
continuously
Draught/air movement
Table/floor vibrations
The weighing plate is in contact with foreign mat-
ters or is installed incorrectly.
Electromagnetic fields/static charging (choose a
different location for the scale, switch off an inter-
fering device if possible)
The weighing result is obvi-
ously incorrect
The scale display is not set to zero
Incorrect adjustment.
Great fluctuations in temperature .
Warm-up time was ignored.
Electromagnetic fields/static charging (choose a
different location for the scale, switch off an inter-
fering device if possible)
Should other error messages occur, switch the scale off and then on again. If the er-