3.7 Connecting to the weighing terminal
Deadweight cell output KERN KFP weighing platform connection
EXC+(5V) See marking of the deadweight cell
EXC-(0)
SIG-
SIG+
4 Operating limits
The weighing platform is designed so robustly that an occasional exceeding of the
maximum weighing load does not lead to damage.
The static bearing capacity, i.e. the maximum permissible load, depends on the type
of load carrying (position A – C). The maximum static bearing capacity may not be
exceeded.
Avoid falling loads, shock loads as well as impacts from the side.
Model A B C
KFP 6V20M, KFP 6V20LM 9 kg 6 kg 3 kg
KFP 15V20M
22 kg
15 kg
7 kg
KFP 30V20M 45 kg 30 kg 15 kg
KFP 60V20M, KFP 60V20LM 90 kg 60 kg 30 kg
KFP150V20M, KFP 150V20LM 225 kg 150 kg 75 kg
KFP300V20M 450 kg 300 kg 150 kg
KFP600V20AM
900 kg 600 kg 300 kg
KFP_V20-BA-e-1420 7