5 How to install the balance
Prepare the installation site for the IEX balance in a professional manner.
Ensure that the installation site is dry, level and firm.
The permissible operating temperate range is from -10°C to +40°C
The floor base for the installation of the platform balance must be firm enough
for supporting the maximum weight allowed for the load plus the platform
balance.
The IEX balance has a display unit connected to the platform.
The display unit is designed for attachment to the table or installation on the
wall with the help of a holder.
The balance should be unpacked outside the hazardous zone.
Remove all transport locking devices after unpacking:
CAUTION!
To prevent electrostatic charge, connect the balance to earth.
The contact slots for the wires are marked with the icon „ “.
Prior to transport, prepare a PE conductor for electrostatic connection to earth of
the display unit.
The PE conductor should end in a loop that is screwed to the display unit with the
help of nuts (M4).
To ensure that the loop is fixed permanently and in contact with the display unit,
use a lock washer.
The PE conductor should be of yellow and green colour and provide a sectional
surface of 4mm².
Transport Securing
8 Doc.No.IEX-BA-e-1412