8 Technical examination
Technical examination by expert staff is required every three
months.
The examination should include the following:
Connection to earth quality: The quality of the fastening on the housing and the
resistance of the connection - max. 100Ω.
The quality of the connection between the platform and the display unit (flexible
steel tube connecting both subassemblies).
Damage and loss of a connection is inacceptable.
The electrical connection between the housing and the platform the maximum
resistance between the weighing pan (put on) should be less than100Ω.
Stability and levelling of the balance.
Doc.No.IEX-BA-e-1412 13