Press
or key to move cursor to System
Set .
Press ENTER key to enter System Set menu. Press
or key to move cursor to the item to be set.
Press ENTER key to directly change, or enter
corresponding changing interface.
Press ESC key to return.
As for the items of Auto Save , Auto Print , Auto
Delete , Auto Trans. , Key Sound and Warn.
Sound , ENTER key can be pressed to selected
On or Off .
If Auto Save is set to On , the current group data can
be saved automatically after measurement is finished and
the average value is displayed.
If Auto Print is set to On , the current data can be
printed out after measurement is finished and the average
value is displayed.
If Auto Delete is set to On , gross error can be
deleted automatically when average times had been
achieved or measurement had been finished in advance
by pressing AVE. key according to 3бrule. If some data
had been cancelled, additional measurements should be
carried out to satisfy the times set.
If Auto Trans. is set to On , the current group data
can be output in text format via RS232 after measuring
finished and average value displayed.
If Key Sound is set to On , buzzer will send out a
short sound with each pressing.
If Warn. Sound is set to On , buzzer will send out a
long sound in case of measured value exceeds the
tolerance limit, data deletion or other things.
If Auto Down is set to On , Power will turn off if
neither measurement nor any key operation is performed
within 5 minutes.
6.9.1 LCD brightness Set
Press
key to increase brightness.
Press
key to reduce brightness.
Press ENTER key to finish change.
Press ESC key to cancel change.
The brighter of brightness, the deeper the colour; The
darker of brightness, the lighter the colour;
6.9.2 Time and Date Set
When in this interface, current time and date will be
displayed on the screen, the format is “mm/dd/yy”.
Press
or key to input the value, Press
or key to move cursor.
Press ENTER key to finish the change, which current
time and date will be replaced by time and date set. Press
ESC key to cancel the change.
6.10 About software
When in the main display interface, press MENU key to
enter the main menu.
Press
or key to move cursor to About
Software .
Press ENTER key to enter About Software .
Information about the tester and embedded software will
be displayed on this interface.
The software version and embedded software identification
are subjected to change due to the upgrading of software
without notifying in advance.
6.11 Calibration of the Tester
The tester and impact device must be calibrated with a
standard Leeb hardness test block (not included in
delivery) prior to the first use, or reusing after a long term
idle.
One time calibration is enough for each type of impact
device which equipped with a main unit; recalibration is not
necessary after the replacement of impact device later.
Press ENTER key as well as
key at the same time to
enter the interface of software calibration
Impact direction should be set to .
Five points should be measured vertically down on the
Leeb hardness test block.