Manual

Overview:
Function
Available
settings
Description
Low
High*
cap
d (low)
d (high)
1 t 2kg 1kg 500g 500g 200g 100g
3 t 10kg 5kg 2kg 2kg 1kg 500g
5 t 10kg 5kg 2kg 2kg 1kg 500g
10 t 20kg 10kg 5kg 5kg 2kg 1kg
F0 di
Modification of readability
AZn 0 0.5 d
AZn 1* 1 d
AZn 2 2 d
F1 az
Automatic zero point
correction
(zero tracking)
AZn 3 4 d
F2 bt
Not documented
F3 sp
Not documented
F4 ip
Internal A/D converter value
F5 ut Not documented
Modifications may only be carried out by a
specialist with competent knowledge.
SLP 0 Standby mode switched off
SLP 1* Standby mode after 5 minutes
SLP 2 Standby mode after 10 minutes
SLP 3 Standby mode after 20 minutes
F6 sl
Standby mode
see chap. 4.3.1
SLP 4 Standby mode after 30 minutes
F7 gv Not documented
F8 ci
Password entry
In „P1- - - “ display increase the flashing number using
, select number using HOLD. Confirm entry by
ZERO button.
F9 CL Adjustment, see chapter 7
* = default setting
HFM-BA-e-1011
22