Instruction Manual
HFB-BA-e-1212
24
Overview:
Function
Available
settings
Description
F0 CAP
Select capacity
30k Max = 30 kg
Modifications may only be carried out by
a specialist with competent knowledge.
60k Max = 60 kg
150k Max = 150 kg
Default setting HFB 150K50
300k Max = 300 kg
Default setting HFB 300K100
600k Max = 600 kg
Default setting HFB 600K200
F1 unit Not documented
F2 res
Select resolution
inC 5* 5
inC 10 10
F3 a2n
Automatic zero point
correction
(zero tracking)
0.5 d 0.5 d
1 d* 1 d
2 d 2 d
4 d 4 d
F4 Flt
Filter
Flt 1 fast
slow
Flt 2*
Flt 3
Flt 4
Flt 5
F5 inP Internal A/D converter value
F6 bk
Display background
illumination
bk Au Background illumination switches on
automatically when loaded or a
button is pressed
bk of Background illumination off
bk on Background illumination on
F7 off
Automatic shutdown
function
of 0
Balance will switch-off automatically
after the set period.
Selectable 0/3/5/15/20 minutes.
of 3
of 5
of 10
of 20
F8 Grv Not documented
F9 CAL Adjustment, see chapt. 7
* = default setting