User guide
Overview of the possible software settings:
(there may be small differences referring to the model):
Navigation Menu
GROSS
NET
0
Press TARE 3sec
SET
End
md-2md-1
15.010.0
6.05.0
3.02.0 3.02.0
1.51.0
0.0020.0050.01
1g
2g5g
0.5g1g2g
0.2g0.5g1g
19200960048002400
OffOn
OffOn
trAdtrAtEtrAnGrAvdivSA_OFF
CAPCAP
Overload setting by
hardware option
Off300240180120
If the TRAD setting is OFF, this must be set to ON before a verification.
Reset into verifiable setting:
Press the
key and keep it pressed until „A_OFF“ appears on the display.
Press
key repeatedly until „TRAD“ appears.
Now press
key, in the display appears “OFF”.
Now press the
key, „ON“ appears.
Then press subsequently
, , .
The balance is now ready for verification.
In order to check the settings, repeat the process.