Owner's manual

ECB-BA-e-1114
16
8 Troubleshooting
The balance should be switched off for a short time following an interruption in the
programme sequence and disconnected from the mains supply. It is then necessary
to repeat the weighing process from the beginning.
Help:
Interruption Possible cause
Weight display is not illuminated.
The balance is not switched on.
The mains supply connection has been
interrupted (mains cable not plugged
in/faulty).
Power supply interrupted. .
The weight display changes continu-
ally
Draught/air movement
Table/floor vibrations
The weighing plate is in contact with for-
eign matter.
Electromagnetic fields / static charging
(choose different location/switch off inter-
fering device if possible)
The weighing result is obviously in-
correct
The balance display is not set to zero
Adjustment is no longer correct.
Great fluctuations in temperature.
Electromagnetic fields / static charging
(choose different location/switch off inter-
fering device if possible)
Switch the balance off if other error messages should appear and then switch on a-
gain. Contact the manufacturer if the error message does not disappear.