17 Instant help
Fault Possible cause
Display is not lit up.
The instrument is not switched on.
The rechargeable battery is empty.
The display does not change
when a sample is being loaded
Sample dish / dish holder is fitted incorrectly.
The weight display changes
constantly / the stability display
does not appear.
Sample dish has contact with wind protection
device or heated cover.
Draught/air movement
Table/floor vibrations
Electromagnetic fields / static charging (choose
different location/switch off interfering instrument
if possible)
Incorrect measuring result
Check adjustment
No resetting to zero before loading the sample
Measurement is taking too
long
Incorrect setting shutoff criterion
Measurement is
not reproducible
Sample is not homogenous
Drying time is too short
Drying temperature too high (e.g. oxidation
sample material, boiling point of sample
exceeded)
Temperature sensor soiled or defective