KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen email: info@kern-sohn.com Phone: +49-[0]7433- 9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com Operating Manual Counting balances KERN CKE/CDS Version 2.
GB KERN CKE/CDS Version 2.6 01/2015 Operating Manual Counting balances Table of Contents 1 2 MODE – MENU ....................................................................................................... 3 Technical Data ....................................................................................................... 4 2.1 2.2 KERN CKE .......................................................................................................................................... 4 KERN CDS .......
1 MODE – MENU How to invoke the mode menu: Turn on balance, press and hold the tare key and press the ON/OFF key. Let go of the tare key. / CDS Factory settings: 9600bd: YES Fast: 3 Keyboard overview / Function No Yes Using the automatic reference optimization (OPT) the counting accuracy is automatically increased when applying up to 100 items.
2 Technical Data 2.1 KERN CKE KERN CKE 360-3 CKE 2000-2 CKE 3600-2 CKE 6K0.02 0,001 g 0,01 g 0,01 g 0,02 g Weighing range (max) 360 g 2.000 g 3.600 g 6.000 g Taring range (subtractive) 360 g 2.000 g 3.600 g 6.
KERN CKE 8K0.05 CKE 16K0.05 CKE 16K0.1 Readability (d) 0,05 g 0,05 g 0,1 g Weighing range (max) 8.000 g 16.000 g 16.000 g Taring range (subtractive) 8.000 g 16.000 g 16.000 g Reproducibility 0,05 g 0,1 g 0,1g Linearity ±0,15 g ±0,25 g ± 0,3 g Smallest piece weight 0,05 g 0,05 g 0,1 g Adjustment point 2/4/5/7/8 kg 5/10/15/16 kg 5/10/15/16 kg Recommended adjustment weight (F1), not added 5 kg + 2 kg 10 kg + 5 kg 10 kg + 5 kg Appropriate for verification max. 80% rel.
KERN CKE 36k0.1 CKE 65k0.2 CKE 65k0.5 0,1 g 0,2 g 0,5 g Weighing range (max) 36.000 g 65.000 65.000 Taring range (subtractive) 36.000 g 65.000 65.000 Reproducibility 0,2 g 0,4 g 0,5 g Linearity ±0,5 g ±1,0 g ± 1,5 g Smallest piece weight 0,1 g 0,2 g 0,5 g Adjustment point 10/20/30/36 kg 20/30/50/60 kg 20/30/50/60 kg Recommended adjustment weight (F1), not added 20 kg + 10 kg 50 kg 50 kg Readability (d) Appropriate for verification max. 80% rel.
2.2 KERN CDS KERN CDS 4K0.02 CDS 15K0.05 CDS 16K0.1 CDS 30K0.1 CDS 30K0.1L Readability (d) 0,02 g 0,05 g 0,1 g 0,1 g 0,1 g Weighing range (max) 4.000 g 15.000 g 16.000 g 30.000 g 30.000 g Taring range (subtractive) 4.000 g 15.000 g 16.000 g 30.000 g 30.
KERN CDS 36K0.2L CDS 60K0.2 CDS 100K0.5 0,2 g 0,2 g 0,5 g Weighing range (max) 36.000 g 60.000 g 100.000 g Taring range (subtractive) 36.000 g 60.000 g 100.000 g Reproducibility 0,2 g 0,4 g 0,5 g Linearity ± ±0,6 g ±1,0 g ±1,5 g Smallest piece weight 0,2 g 10/15/20/30/36 20/30/50/60 20/50/100 20 kg + 10 kg 50 kg 50 kg + 50 kg Readability (d) Adjustment point kg Recommended adjustment weight (F1), not added Appropriate for verification max. 80% rel.
3 Basic Information (General) It is absolutely necessary that you read and understand the operating instructions prior to installation and commissioning and follow the instructions during the process! 3.1 Proper use The balance you purchased is intended to determine the weighing value of material to be weighed. It is intended to be used as a “non-automatic“ balance, i.e. the material to be weighed is manually and carefully placed in the centre of the weighing plate.
4 Basic Safety Precautions 4.1 Pay attention to the instructions in the Operation Manual Carefully read this operation manual before setup and commissioning, even if you are already familiar with KERN balances. Versions in other languages are non-binding translations. The only binding version is the original document in German. 4.2 Personnel training The appliance may only be operated and maintained by trained personnel. 5 Transport and storage 5.
6 Unpacking, Setup and Commissioning 6.1 Installation Site, Location of Use The balances are designed in a way that reliable weighing results are achieved in common conditions of use. You will work accurately and fast, if you select the right location for your balance.
6.4 Battery operation CKE To insert the batteries (6 x 1.5 V) remove the battery compartment cover. Remove it with the help of a coin. In the each battery tube insert three batteries in the same polarity sense. Screw down again the battery cover. To save the battery, the background illumination can be switched off (see chap. 1 Mode Menu). Moreover the AUTO-OFF function can be activated (see chap. 1 Mode Menu).
6.7 Adjustment As the acceleration value due to gravity is not the same at every location on earth, each balance must be coordinated - in compliance with the underlying physical weighing principle - to the existing acceleration due to gravity at its place of location (only if the balance has not already been adjusted to the location in the factory). This adjustment process must be carried out during the initial start-up, after change in location and variation of surrounding temperature.
7 Operation 7.1 Control panel CKE/CDS ON/OFF Taring; Invoke the mode menu by pressing the ON/OFF key. Generate reference using 5 parts; In mode menu: NO function Generate reference using 10 parts Generate reference using 20 parts; In mode menu: YES function Switchover pcs ↔ g; In menu: Mode function. Print out weighing result. Display icon Significance ==OVERLOAD== Overload: Weighing range exceeded ============= Underload: Weighing range not achieved << 14 .
7.2 Operation 7.2.1 Adjusting 7.2.2 Speed 7.2.
CKE/CDS-BA-e-1526
7.2.4 Display background illumination Switch on the balance and the zero display, then call-up the balance menu as specified in chapter 1. Select menu item „Backlight“ using the button necessary for the respective model. Acknowledge by pressing the „YES“ button in order to switch on the background illumination permanently. By pressing the „NO“ button, the background illumination is switched off.
7.3 Data output RS 232 C Technical Data 8-bit ASCII Code 1 start bit, 8 data bits, 1 stop bit, no parity bit Baud rate selectable from 2400, 4800, 9600 Baud (factory setting) and 19200 Baud. Sub-D plug 9-channel required For operation with interface faultless operation is only ensured with the correct KERN – interface cable (max.
7.4 Interface RS 232C Data output via interface RS 232C General Information The previous condition for the data transfer between balance and a peripherical device (e.g. printer, PC ...) is that the appliances are set to the same interface parameters (e.g. baud rate, parity ...). 7.4.1 There are 4 kinds of data output via RS 232C Data output using the PRINT key The printing process can be triggered by pressing the PRINT key. The settings AUTOPRINT and AUTOPRINT PC should be disabled for this process.
7.5 Printer The serial interface RS 232 facilitates the connection of a printer. The printout shows the weight in grams. In counting mode either the piece number or the weight details will be printed out. In percentage mode the percentage proportion or the weight details will be printed out. Printout will take place after pressing the PRINT key. It is possible to number each printout continuously with the help of the numerator.
8 Service, maintenance, disposal 8.1 Cleaning Before cleaning, please disconnect the appliance from the operating voltage. Please do not use aggressive cleaning agents (solvents or similar agents), but a cloth dampened with mild soap suds. Ensure that no liquid penetrates into the device and wipe with a dry soft cloth. Loose residue sample/powder can be removed carefully with a brush or manual vacuum cleaner. Spilled weighing goods must be removed immediately. 8.
10 Declaration of conformity KERN & Sohn GmbH D-72322 Balingen-Frommern Postfach 4052 E-Mail: info@kern-sohn.de Tel: 0049-[0]7433- 9933-0 Fax: 0049-[0]7433-9933-149 Internet: www.kern-sohn.