Manual

CFS/CCS-BA-e-1220 33
„LoAd 3“ will be displayed
50 kg adjustment weight placed carefully in the
centre of the weighing plate, wait for stability display
and confirm with .
After the adjustment the balance will carry out a self-
test. Remove adjusting weight during selftest,
balance will return into weighing mode automatically.
In case of fault or a wrong weight an error message
( ) will appear, then repeat the
linearization procedure.
8.2.1 Table linearisation points CFS 50K-3 models.
Max: 50 kg
Display
Weight
„L0Ad 0“
0 kg
„L0Ad 1“
15 kg
„L0Ad 2“
30 kg
„L0Ad 3“
50 kg