User guide
17.1.1 Setting an adjustment report
The following settings can be enabled (printout takes place) or disabled (printout
blocked) by selecting YES or NO.
Sample adjustment record:
Select P 5.1 and adjust the desired settings.
Sample adjustment record:
P5.1.1 DRAFT
Enter a project name, maximum 16 characters
P5.1.2 KIND JUST YES
Print kind of adjustment
P5.1.3 USER YES
Print name of user
P5.1.4 DRAFT YES
Print project name (See. P5.1.1)
P5.1.5 DATE YES
Print date of adjustment
P5.1.6 TIME YES
Print time of adjustment
P5.1.7 ID WEIGHING
SCALE
YES
Print serial number of weighing scale
P5.1.8 DIFF.JUST. YES
Print difference between currently adjustment
weight and previous adjustment weight.
P5.1.9
LINES
YES
Print line between date and signature
P5.1.10 SIGNATURE YES
Print box for signature of implementing person
Printout example adjustment report:
-------------------Just.Prot.------------------
Kind Just External
User Kern
Draft
Datum 21.11.2013
Time 9:54:39
ID Balance 132051
Diff.Just. -0.2 g
Signature
……………………………………………..
98 AEJ-C/AES-C/PLJ-C-BA-e-1412