User guide

16.4.2 Remote control instructions
Feedback by weighing scale after sending remote control command:
XX_A CR LF
Instruction accepted; will be executed
XX_I CR LF
Instruction received; impossible to carry out
XX_^ CR LF
Command accepted but max range exceeded
XX_v CR LF
Command accepted but min range exceeded
XX_E CR LF
Error during execution, timeout for stable weighing value
exceeded
XX
Name of command
1 – 3
4
5
6
7
8 – 16
17
18 - 20
21
22
Command
Blank
Stability
symbol
Blank
Character
Weight
Blank
Unit
CR
LF
Instruction: 1-3
Character
Stabilisation
symbol
[blank space]
If stable
[?]
If stable
[^]
If error (Command + exceeded)
[v]
If error (Command exceeded)
Character
[blank space]
For positive and negative values
Weight
9 digits
With right justifications
Unit
3 digits
With left justification
94 AEJ-C/AES-C/PLJ-C-BA-e-1412