User guide

7.2 Menu settings
Menu item „P1 ADJUSTMENT“ [Adjustment]
AEJ-C / PLJ-C AES-C
P1.1 INT.JUST. -
Internal adjustment, ILJ-C only
- P1.2 EXT.JUST.
External adjustment *
- P1.3 BE.JUST.
User adjustment
P1.4 TEST JUST. -
Adjustment test
* = Locked in weighing scales with type approval certificate [Er 9 lock].
7.2.1 P1.1 INT.J UST. - internal adjustment (only models AE J -C and PLJ -C)
For automatic internal adjustment see chap. 7.1
Starting internal adjustment manually
Press the CAL-key or invoke menu “P1.1 INT.JUST.“
Press
or
Internal adjustment will be taking place.
Ensure that there are no objects on the weighing pan.
You can cancel the adjustment by pressing the UNIT-key.
After successful adjustment the weighing scale will return automatically to
the menu item. To return to weighing mode, press UNIT.
AEJ-C/AES-C/PLJ-C-BA-e-1412 27