Manual
80 AEJ_N/AES_N-BA-e-1321
14.5.2 PC controlled output
Response message of balance after remote instruction was sent:
XX_ Instruction:
XX_A CR LF Instruction accepted; will be executed
XX_ I CR LF Instruction received; impossible to carry out
XX_^ CR LF Instruction received but time overflow error occurred
XX_v CR LF Instruction received, but insufficient load
XX_E CR LF Error during execution, timeout for stable weighing value
exceeded
Data record format:
1 - 3
4
5
6
7
8 - 16
17
18 - 20
21
22
Remote control
command
Blank
Stability display
Blank
Signs
Weight value
Blank
Unit
CR
LF
Instruction:
1. up to 3 signs
Stability display
Blank character for stable value
? For instable value
^ for excess load
v for underload
Signs
Blank character for positive value
Minus sign for negative value
Weight value
9 signs, right justified
Unit
3 signs, left justified
14.5.3 Output of date/time
The display of date and time can be enabled under menu item “P2 GLP”
(see chap. 12):
• PdAt – yes
• Ptin - yes