AEJ_N/AES_N-BA-e-1321 79
14.5 Output log
14.5.1 Issue after pressing the PRINT ´button
Stable or instable weighing values are sent according to menu settings “P4.5” PStb
(see chap. 14.3)
Data record format:
1
2
3
4 - 12
13
14 - 16
17
18
Stability display
Blank
Signs
Weight value
r Blank
Unit
CR
LF
Stability display
Blank character for stable value
? For instable value
^ for excess load
v for underload
Signs
Blank character for positive value
Minus sign for negative value
Weight value
9 signs, right justified
Unit
3 signs, left justified